CONCRETE SURFACE RESISTIVITY : Giatec Surf™ is a laboratory test device for rapid, easy and accurate measurement
of the surface electrical resistivity of concrete based on the four-probe (Wenner-Array) technique. The Surf™ patented technology automatically measures resistivity around the concrete specimen using four channels of 4-probe array (located at 90° from each other). The PC software generates the required reports according to the standard specifications.
Surf™ can be used to determine the chloride permeability of concrete in accordance with AASHTO TP95 and the upcoming ASTM standard. The measurement data can be used for durability-based quality control of concrete as well as the service life design of concrete structures
APPLICATIONS
The electrical resistivity of concrete is correlated well with important durability parameters such as permeability, diffusivity and in general the micro-structure characteristics of concrete. It is a fast and easy method of quality control during new construction.
Surf™ is a unique device for automatic 4-channel resistivity measurement suitable for investigating the micro-structural properties of concrete including:
- Durability-based quality control of concrete
- Diffusion of chloride in concrete
- Setting time of fresh concrete
- Crack detection in concrete
- Water content in fresh concrete
FEATURES
- Patented technology
- The only device in the market fully compliant with both AASHTO T95 and the upcoming ASTM standard
- Easy to use (see demo on Youtube)
- Variable frequency (13 – 100 Hz)
- Wide resistivity range (0.1 – 1,000 kΩ.cm)
- Fast measurement (8 measurements < 15s)
- Four-channel four-probe measurement
- Limiting moisture loss
- Automatic report generation with PC software
- Fresh concrete testing/crack detection applications
- Continuous measurement mode
- USB connection to computer
- CORRELATED WELL WITH IMPORTANT DURABILITY PARAMETERS
- INVESTIGATING THE MICRO-STRUCTURAL PROPERTIES OF CONCRETE
- AUTOMATIC 4-CHANNEL RESISTIVITY MEASUREMENT
- FAST AND EASY METHOD OF QUALITY CONTROL
- WIDE RESISTIVITY RANGE
SPECIFICATION
GENERAL
Reading range | Frequency range | Accuracy |
0.1-100KΩ.cm | 13 – 100 Hz | ± (0.1+1%) |
100-1000KΩ.cm | 13 – 100 Hz | ± (1+1%) |
MEASUREMENT TIME
Frequency | Single measurement time | Testing time (8 measurements) |
13-100 Hz | 1.5 seconds | <15 seconds |
OPERATING CONDITIONS
Type | Value |
Operating temperature | 15 ~ 45 °C |
Operating humidity | 30 ~ 80% |
Storage temperature | 0 ~ 60°C |
Operating voltage/ current | 100-250 V ± 10%, 60Hz |